JEM-ARM200F Opens Doors to New Ultra Micro Analysis

JEOL JEM-ARM200FThe JEM-ARM200F, incorporating a spherical aberration corrector for electron optic system as standard, has achieved a scanning transmission image (STEM-HAADF) resolution of 0.08 nm, the highest in the world among the commercial transmission electron microscopes.

  • World's highest STEM (HAADF) resolution of 0.08 nm guaranteed
    The ARM200F, incorporating a spherical aberration corrector for electron optic system as standard and the maximum level of electrical and mechanical stability, has achieved a scanning transmission image (STEM-HAADF) resolution of 0.08 nm, the highest in the world among the commercial transmission electron microscopes.The electron probe, after its aberrations are corrected, features a current density level higher by an order of magnitude than conventional transmission electron microscopes.With this probe finely focused, the ARM200F is capable of atomic level analysis, substantially reducing measurement time and imoproving throughput.
     
  • Enhanced electrical stability
    To achieve an atomic level resolution, the power supply to control the electron optic system needs to be stabilized.The ARM 200F has reduced the fluctuation of high voltage and objective current by 50% of conventional transmission electron microscopes, significantly enhancing its electrical stability.
     
  • Enhanced mechanical stability
    To enable atomic level analysis and imaging utilizing spherical aberration correctors for electron optic and image forming systems, vibration and distortion on the atomic level need to be controlled.The ARM 200F features an overall mechanical strength twice higher than conventional transmission microscopes by enlarging the coolumn size to improve the rigidity and optimizing the console structure, enhancing mechanical stability.
     
  • Wide ranging STEM analytical capabilities
    Two types of dark area detectors with different STEM detection angles (one as standard), bright area detector (standard), and backscattered electron detector (option) can be simultaneously installed.A new scanning image acquisigion system is capable of acquiring 4 different types of signals, enabling simultaneous observation of these 4 images.
     
  • Environmental measures
    Atomic level vibration and distortion are also caused by changes in temperature and stray magnetic field in the instrument room.The ARM 200F features a set of thermal and magnetic shields as standard.The column is also housed in a cover to prevent the column surface temperature from fluctuating as the ambient convection changes.
     
  • Optional spherical aberration corrector for image forming system
    With an optional spherical aberration corrector, the resolution of transmission electron microscope image (TEM) can be improved to 0.11 nm.

JEM-ARM200F Opens Doors to New Ultra Micro Analysis

JEOL JEM-ARM200FThe JEM-ARM200F, incorporating a spherical aberration corrector for electron optic system as standard, has achieved a scanning transmission image (STEM-HAADF) resolution of 0.08 nm, the highest in the world among the commercial transmission electron microscopes.

  • World's highest STEM (HAADF) resolution of 0.08 nm guaranteed
    The ARM200F, incorporating a spherical aberration corrector for electron optic system as standard and the maximum level of electrical and mechanical stability, has achieved a scanning transmission image (STEM-HAADF) resolution of 0.08 nm, the highest in the world among the commercial transmission electron microscopes.The electron probe, after its aberrations are corrected, features a current density level higher by an order of magnitude than conventional transmission electron microscopes.With this probe finely focused, the ARM200F is capable of atomic level analysis, substantially reducing measurement time and imoproving throughput.
     
  • Enhanced electrical stability
    To achieve an atomic level resolution, the power supply to control the electron optic system needs to be stabilized.The ARM 200F has reduced the fluctuation of high voltage and objective current by 50% of conventional transmission electron microscopes, significantly enhancing its electrical stability.
     
  • Enhanced mechanical stability
    To enable atomic level analysis and imaging utilizing spherical aberration correctors for electron optic and image forming systems, vibration and distortion on the atomic level need to be controlled.The ARM 200F features an overall mechanical strength twice higher than conventional transmission microscopes by enlarging the coolumn size to improve the rigidity and optimizing the console structure, enhancing mechanical stability.
     
  • Wide ranging STEM analytical capabilities
    Two types of dark area detectors with different STEM detection angles (one as standard), bright area detector (standard), and backscattered electron detector (option) can be simultaneously installed.A new scanning image acquisigion system is capable of acquiring 4 different types of signals, enabling simultaneous observation of these 4 images.
     
  • Environmental measures
    Atomic level vibration and distortion are also caused by changes in temperature and stray magnetic field in the instrument room.The ARM 200F features a set of thermal and magnetic shields as standard.The column is also housed in a cover to prevent the column surface temperature from fluctuating as the ambient convection changes.
     
  • Optional spherical aberration corrector for image forming system
    With an optional spherical aberration corrector, the resolution of transmission electron microscope image (TEM) can be improved to 0.11 nm.

View the JEM-ARM200F slide show here
 

JEM-ARM200F Opens Doors to New Ultra Micro Analysis

JEOL JEM-ARM200FThe JEM-ARM200F, incorporating a spherical aberration corrector for electron optic system as standard, has achieved a scanning transmission image (STEM-HAADF) resolution of 0.08 nm, the highest in the world among the commercial transmission electron microscopes.

  • World's highest STEM (HAADF) resolution of 0.08 nm guaranteed
    The ARM200F, incorporating a spherical aberration corrector for electron optic system as standard and the maximum level of electrical and mechanical stability, has achieved a scanning transmission image (STEM-HAADF) resolution of 0.08 nm, the highest in the world among the commercial transmission electron microscopes.The electron probe, after its aberrations are corrected, features a current density level higher by an order of magnitude than conventional transmission electron microscopes.With this probe finely focused, the ARM200F is capable of atomic level analysis, substantially reducing measurement time and imoproving throughput.
     
  • Enhanced electrical stability
    To achieve an atomic level resolution, the power supply to control the electron optic system needs to be stabilized.The ARM 200F has reduced the fluctuation of high voltage and objective current by 50% of conventional transmission electron microscopes, significantly enhancing its electrical stability.
     
  • Enhanced mechanical stability
    To enable atomic level analysis and imaging utilizing spherical aberration correctors for electron optic and image forming systems, vibration and distortion on the atomic level need to be controlled.The ARM 200F features an overall mechanical strength twice higher than conventional transmission microscopes by enlarging the coolumn size to improve the rigidity and optimizing the console structure, enhancing mechanical stability.
     
  • Wide ranging STEM analytical capabilities
    Two types of dark area detectors with different STEM detection angles (one as standard), bright area detector (standard), and backscattered electron detector (option) can be simultaneously installed.A new scanning image acquisigion system is capable of acquiring 4 different types of signals, enabling simultaneous observation of these 4 images.
     
  • Environmental measures
    Atomic level vibration and distortion are also caused by changes in temperature and stray magnetic field in the instrument room.The ARM 200F features a set of thermal and magnetic shields as standard.The column is also housed in a cover to prevent the column surface temperature from fluctuating as the ambient convection changes.
     
  • Optional spherical aberration corrector for image forming system
    With an optional spherical aberration corrector, the resolution of transmission electron microscope image (TEM) can be improved to 0.11 nm.

View the JEM-ARM200F slide show here
 

JEM-ARM200F Specifications

Resolution
Scanning transmission image 1) 0.08 nm 2) (at 200kV)
Transmission image
Point image
Lattice image
 
0.19 nm (at 200kV), 0.11 nm with TEM Cs corrector 3) (at 200kV)
0.10 nm
Magnification
Scanning transmission image 100 to 150,000,000x
Transmission image

50 to 2,000,000x

Electron gun Schottky field emission gun
Accelerating voltage 80 to 200 kV 4)
Specimen Stage
Stage Eucentric side entry goniometer stage
Specimen size 3 mm dia.
Tilt angle Up to 25° (with double tilt holder)
Movement range X/Y: ±1.0 mm (motor drive/piezo drive)
Cs Correctors
STEM Cs corrector Standard
TEM Cs corrector Optional
Optional accessories Energy Dispersive X-Ray Spectrometer (EDS)
Electron energy-loss spectrometer (EELS)
CCD camera, etc.

*Dark field ring detector used

JEM-ARM200F Specifications

Resolution
Scanning transmission image1) 0.08 nm 2) (at 200kV)
Transmission image
Point image
Lattice image
 
0.19 nm (at 200kV), 0.11 nm with TEM Cs corrector 3) (at 200kV)
0.10 nm
Magnification
Scanning transmission image 100 to 150,000,000x
Transmission image

50 to 2,000,000x

Electron gun Schottky field emission gun
Accelerating voltage 80 to 200 kV 4)
Specimen Stage
Stage Eucentric side entry goniometer stage
Specimen size 3 mm dia.
Tilt angle Up to 25° (with double tilt holder)
Movement range X/Y: ±1.0 mm (motor drive/piezo drive)
Cs Correctors
STEM Cs corrector Standard
TEM Cs corrector Optional
Optional accessories Energy Dispersive X-Ray Spectrometer (EDS)
Electron energy-loss spectrometer (EELS)
CCD camera, etc.

*Dark field ring detector used

JEM-ARM200F Specifications

Resolution
Scanning transmission image 1) 0.08 nm 2) (at 200kV)
Transmission image
Point image
Lattice image
 
0.19 nm (at 200kV), 0.11 nm with TEM Cs corrector 3) (at 200kV)
0.10 nm
Magnification
Scanning transmission image 100 to 150,000,000x
Transmission image

50 to 2,000,000x

Electron gun Schottky field emission gun
Accelerating voltage 80 to 200 kV 4)
Specimen Stage
Stage Eucentric side entry goniometer stage
Specimen size 3 mm dia.
Tilt angle Up to 25° (with double tilt holder)
Movement range X/Y: ±1.0 mm (motor drive/piezo drive)
Cs Correctors
STEM Cs corrector Standard
TEM Cs corrector Optional
Optional accessories Energy Dispersive X-Ray Spectrometer (EDS)
Electron energy-loss spectrometer (EELS)
CCD camera, etc.

1) With HAADF (high-angle annular dark-field) detector
2) Verified using Ge(112) specimen
3) Optional
4) Standard voltage: 120 kV, 200 kV

JEM-ARM200F Specifications

Resolution
Scanning transmission image 1) 0.08 nm 2) (at 200kV)
Transmission image
Point image

Lattice image
 
0.19 nm (at 200kV)
0.11 nm with TEM Cs corrector 3) (at 200kV)
0.10 nm
Magnification
Scanning transmission image 100 to 150,000,000x
Transmission image

50 to 2,000,000x

Electron gun Schottky field emission gun
Accelerating voltage 80 to 200 kV 4)
Specimen Stage
Stage Eucentric side entry goniometer stage
Specimen size 3 mm dia.
Tilt angle Up to 25° (with double tilt holder)
Movement range X/Y: ±1.0 mm (motor drive/piezo drive)
Cs Correctors
STEM Cs corrector Standard
TEM Cs corrector Optional
Optional accessories Energy Dispersive X-Ray Spectrometer (EDS)
Electron energy-loss spectrometer (EELS)
CCD camera, etc.

1) With HAADF (high-angle annular dark-field) detector
2) Verified using Ge(112) specimen
3) Optional
4) Standard voltage: 120 kV, 200 kV

JEM-ARM200F Specifications

Resolution
Scanning transmission image 1) 0.08 nm 2) (at 200kV)
Transmission image
Point image

Lattice image
 
0.19 nm (at 200kV)
0.11 nm with TEM Cs corrector 3) (at 200kV)
0.10 nm
Magnification
Scanning transmission image 200 to 150,000,000x
Transmission image

50 to 2,000,000x

Electron gun Schottky field emission gun
Accelerating voltage 80 to 200 kV 4)
Specimen Stage
Stage Eucentric side entry goniometer stage
Specimen size 3 mm dia.
Tilt angle Up to 25° (with double tilt holder)
Movement range X/Y: ±1.0 mm (motor drive/piezo drive)
Cs Correctors
STEM Cs corrector Standard
TEM Cs corrector Optional
Optional accessories Energy Dispersive X-Ray Spectrometer (EDS)
Electron energy-loss spectrometer (EELS)
CCD camera, etc.

1) With HAADF (high-angle annular dark-field) detector
2) Verified using Ge(112) specimen
3) Optional
4) Standard voltage: 120 kV, 200 kV

Please Note